Part Number Hot Search : 
NJ1800DL BU1706A 2SK786 AA88048 11110 FF4N60 T2301 C908A
Product Description
Full Text Search
 

To Download 5990-6583EN Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
  keysight technologies m9290a cxa-m pxie x-series signal analyzer 10 hz to 3.0, 7.5, 13.6 or 26.5 ghz data sheet
table of contents technical specifications ................................................................... 3 definitions and conditions ........................................................ 3 recommended best practices in use ...................................... 3 frequency and time specifications ......................................... 4 amplitude accuracy and range specifications ...................... 6 dynamic range specifications ................................................ 9 powersuite measurement specifications .............................. 12 tracking generator specifications ........................................ 13 general specifications ........................................................... 14 inputs and outputs ................................................................. 14 i/q analyzer ............................................................................ 15 system requirements ............................................................ 15 software ......................................................................................... 16 related literature ........................................................................... 17 deploy a smaller microwave footprint in test system development, one of your crucial requirements is doing more in less spacebut this often means tradeoffs between footprint and precision in signal analysis. that is, until now: weve expanded the keysight x-series with the cxa-m, a pxie signal analyzer that offers fully specified performance up to 26.5 ghz. it lets you handle rf and microwave signals in four slots, and you can leverage your existing code. choose the cxa-m and deploy a smaller microwave footprint.
03 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet technical specifcations defnitions and conditions temperatures referred to in this document are defined as follows: C full temperature range = individual module temperature of 5 to 68 c, as reported by the module, and environment temperature of 0 to 55 c. C controlled temperature range = individual module temperature of 25 to 40 c, as reported by the module, and environment temperature of 20 to 30 c. specifications describe the warranted performance of calibrated instruments. data represented in this document are specifications under the following conditions unless otherwise noted. C it is within its calibration cycle C under auto couple control, except when auto sweep time rules = accy C the analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on; if it had previously been stored at a temperature range inside the allowed storage range, but outside the allowed operating range C the analyzer has been turned on at least 30 minutes with auto align set to normal, or, if auto align is set to off or partial, alignments must have been run recently enough to prevent an alert message; if the alert condition is changed from time and temperature to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user 95th percentile values indicate the breadth of the population (approx. 2) of perfor - mance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 c. in addition to the statistical observations of a sample of instruments, these values include the effects of the uncertainties of external calibration references. these values are not warranted. these values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. typical describes additional product performance information that is not covered by the product warranty. it is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 c. typical performance does not include measurement uncertainty. nominal values indicate expected performance, or describe product performance that is useful in the application of the product, but are not covered by the product warranty. recommended best practices in use C use slot blockers and emc filler panels in empty module slots to ensure proper operating temperatures. keysight chassis and slot blockers optimize module temperature performance and reliability of test. C set chassis fan to high at environmental temperatures above 45c. get more information this cxa-m pxie x-series signal analyzer data sheet is a summary of the complete specifications and conditions available in the cxa-m pxie signal analyzer specification guide. the cxa-m pxie signal analyzer specifica - tion guide can be obtained on the web at: www.keysight.com/find/ cxa-m_manuals for ordering information, refer to the cxa-m pxie signal analyzer configuration guide (5992-0193en).
04 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet frequency and time specifcations frequency range dc coupled ac coupled option f03 10 hz to 3.0 ghz 10 mhz to 3.0 ghz option f07 10 hz to 7.5 ghz 10 mhz to 7.5 ghz option f13 10 hz to 13.6 ghz 10 mhz to 13.6 ghz option f26 10 hz to 26.5 ghz 10 mhz to 26.5 ghz band lo multiple (n) 0 1 10 hz to 3.08 ghz 1 2 2.95 to 7.575 ghz 2 2 7.45 to 9.55 ghz 3 2 9.45 to 12.60 ghz 4 2 12.50 to 13.05 ghz 4 4 12.95 to 13.80 ghz 5 4 13.40 to 15.55 ghz 6 4 15.45 to 19.35 ghz 7 4 19.25 to 21.05 ghz 8 4 20.95 to 22.85 ghz 9 4 22.75 to 24.25 ghz 10 4 24.15 to 26.55 ghz frequency reference accuracy [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] aging rate option pfr standard 1 x 10 -7 / year 1 x 10 -6 / year 1.5 x 10 -7 / 2 years temperature stability option pfr standard 20 to 30 c 1.5 x 10 -8 2 x 10 -6 full temperature range 5 x 10 -8 2 x 10 -6 achievable initial calibration accuracy option pfr standard 4 x 10 -8 1.4 x 10 -6 example frequency reference accuracy (with option pfr) = (1 x 1 x 10 -7 + 5 x 10 -8 + 4 x 10 -8 ) 1 year after last adjustment = 1.9 x 10 -7 residual fm option pfr 0.25 hz p-p in 20 ms nominal standard 10 hz p-p in 20 ms nominal frequency readout accuracy (start, stop, center, marker) (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x rbw + 2 hz + 0.5 x horizontal resolution 1 ) marker frequency counter accuracy (marker frequency x frequency reference accuracy + 0.100 hz) delta counter accuracy (delta frequency x frequency reference accuracy + 0.141 hz) counter resolution 0.001 hz frequency span (fft and swept mode) range 0 hz (zero span), 10 hz to maximum frequency of instrument resolution 2 hz accuracy swept (0.25 % x span + horizontal resolution) fft (0.10 % x span + horizontal resolution) 1. horizontal resolution is span/(sweep points C 1)
05 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet sweep time and triggering range span = 0 hz 1 s to 6000 s span 10 hz 1 ms to 4000 s accuracy span 10 hz, swept 0.01 % nominal span 10 hz, fft 40 % nominal span = 0 hz 1 % nominal trigger free run, video, external, rf burst, periodic timer trigger delay span = 0 hz or fft C150 to +500 ms span 10 hz, swept 1 s to 500 ms resolution 0.1 s time gating gate methods gated lo; gated video; gated fft gate length range (except method = fft) 100.0 ns to 5.0 s gate delay range 0 to 100.0 s gate delay jitter 33.3 ns p-p nominal sweep (trace) point range all spans 1 to 40001 resolution bandwidth (rbw) range (C3.01 db bandwidth) 1 hz to 3 mhz (10 % steps), 4, 5, 6, 8 mhz bandwidth accuracy (power) 1 hz to 750 khz 1.0 % ( 0.044 db) nominal 820 khz to 1.2 mhz 2.0 % ( 0.088 db) nominal 1.3 to 2.0 mhz 0.13 db nominal 2.2 to 3 mhz 0.22 db nominal 4 to 8 mhz 0.32 db nominal bandwidth accuracy (C3.01 db) 1 hz to 1.3 mhz 2 % nominal rbw range selectivity (C60 db/C3 db) 4.1:1 nominal emi bandwidth (cispr compliant) 200 hz, 9 khz, 120 khz, 1 mhz (option emc) emi bandwidth (mil std 461e compliant) 10 hz, 100 hz, 1 khz, 10 khz, 100 khz, 1 mhz (option emc) analysis bandwidth 1 maximum bandwidth option b25 25 mhz standard 10 mhz video bandwidth (vbw) range 1 hz to 3 mhz (10 % steps), 4, 5, 6, 8 mhz, and wide open (labeled 50 mhz) accuracy 6 % nominal measurement speed 2 local measurement and display update rate 11 ms (90/s) nominal remote measurement and lan transfer rate 6 ms (167/s) nominal marker peak search 5 ms nominal center frequency tune and transfer 22 ms nominal measurement/mode switching 75 ms nominal 1. analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency, or modulation domain. 2. sweep points = 101. frequency and time specifcations (continued)
06 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet amplitude accuracy and range specifcations amplitude range measurement range preamp off displayed average noise level (danl) to +30 dbm preamp on displayed average noise level (danl) to +23 dbm input attenuator range standard 0 to 70 db in 10 db steps option fsa 0 to 70 db in 2 db steps, 7.5 ghz 0 to 70 db in 10 db steps, 7.5 to 26.5 ghz maximum safe input level average total power +30 dbm (1 w) input attenuation 10 db, preamp off +30 dbm (1 w) input attenuation 20 db, preamp on peak pulse power +50 dbm (100 w) < 10 s pulse width, < 1 % duty cycle, and input attenuation 30 db dc volts ac coupled 50 vdc dc coupled 0.2 vdc display range log scale 0.1 to 1 db/division in 0.1 db steps 1 to 20 db/division in 1 db steps (10 display divisions) linear scale 10 divisions scale units dbm, dbmv, dbv, dbma, dba, v, w, a frequency response specifcation 95th percentile ( 2 ) (10 db input attenuation, 20 to 30 c, = nominal standard deviation) 9 khz to 10 mhz 0.50 db 0.4 db 10 mhz to 3 ghz 0.65 db 0.5 db 3 to 13.6 ghz 1.30 db 0.8 db 13.6 to 19.3 ghz 1.50 db 1.0 db 19.3 to 24.2 ghz 2.20 db 1.3 db 24.2 to 26.5 ghz 2.50 db 1.3 db preamp on (option p03, p07, p13, p26) 100 khz to 10 mhz 0.5 db 10 mhz to 3 ghz 1.0 db 3 to 7.5 ghz 1.2 db 7.5 to 13.6 ghz 1.0 db 13.6 to 21 ghz 1.2 db 21 to 24.2 ghz 1.8 db 24.2 to 26.5 ghz 2.4 db input attenuation switching uncertainty specifcations additional information attenuation > 2 db, preamp off 50 mhz (reference frequency) 0.3 db 0.15 db typical relative to 10 db (reference setting) 100 khz to 3.0 ghz 0.30 db nominal 3.0 to 7.5 ghz 0.50 db nominal 7.5 to 26.5 ghz 0.70 db nominal
07 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet total absolute amplitude accuracy (10 db attenuation, 20 to 30 c, 1 hz rbw 1 mhz, input signal C10 to C50 dbm, all settings auto-coupled except auto swp time = accy, any reference level, any scale, = nominal standard deviation) at 50 mhz 0.40 db at all frequencies (0.40 db + frequency response) 100 khz to 3 ghz 0.60 db (95th percentile 2) preamp on (option p03/p07/p13/p26) (0.36 db + frequency response) (95th percentile) precision amplitude accuracy (option paa) preamp off (10 db attenuation) specifcation additional information 9 khz to 10 mhz 0.70 db 0.31 db typical 10 mhz to 3 ghz 0.85 db 0.40 db typical 3 to 7.5 ghz 0.95 db 0.44 db typical 7.5 to 9.55 ghz 0.90 db 0.38 db typical 9.55 to 12.55 ghz 0.95 db 0.44 db typical 12.55 to 13.6 ghz 0.90 db 0.32 db typical 13.6 to 15.55 ghz 0.95 db 0.35 db typical 15.55 to 22.8 ghz 1.10 db 0.50 db typical 22.8 to 24.2 ghz 1.20 db 0.50 db typical 24.2 to 26.5 ghz 1.30 db 0.57 db typical preamp on (0 db attenuation) 100 khz to 10 mhz 0.30 db typical 10 mhz to 3 ghz 0.65 db typical 3 to 7.5 ghz 0.73 db typical 7.5 to 9.55 ghz 0.49 db typical 9.55 to 12.55 ghz 0.71 db typical 12.55 to 19.3 ghz 0.65 db typical 19.3 to 21 ghz 0.80 db typical 21 to 22.8 ghz 1.20 db typical 22.8 to 24.2 ghz 1.39 db typical 24.2 to 26.5 ghz 1.66 db typical input voltage standing wave ratio (vswr) ( 10 db attenuation) 10 mhz to 3 ghz < 1.2 nominal 3 to 7.5 ghz < 1.4 nominal 7.5 to 13.6 ghz < 1.6 nominal 13.6 to 24.2 ghz < 1.8 nominal 24.2 to 26.5 ghz < 2.2 nominal resolution bandwidth switching uncertainty (referenced to 30 khz rbw) 1 hz to 3 mhz rbw 0.15 db 4, 5, 6, 8 mhz rbw 1.0 db reference level range log scale C170 to +23 dbm in 0.01 db steps linear scale same as log (707 pv to 3.16 v) accuracy 0 db display scale switching uncertainty switching between linear and log 0 db log scale/div switching 0 db amplitude accuracy and range specifcations (continued)
08 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet display scale fdelity C80 dbm input mixer level < C10 dbm 0.15 db total trace detectors normal, peak, sample, negative peak, log power average, rms average, and voltage average preamplifer (option p03/p07/p13/p26) frequency range option p03 100 khz to 3.0 ghz option p07 100 khz to 7.5 ghz option p13 100 khz to 13.6 ghz option p26 100 khz to 26.5 ghz gain 100 khz to 26.5 ghz +20 db nominal noise fgure 10 mhz to 3 ghz 10 db nominal 3 to 26.5 ghz danl + 176.24 db nominal amplitude accuracy and range specifcations (continued)
09 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet 1 db gain compression (two-tone) total power at input mixer preamp off 10 mhz to 7.5 ghz +6 dbm nominal 7.5 to 26.5 ghz +4 dbm nominal preamp on 10 mhz to 7.5 ghz C15 dbm nominal 7.5 to 26.5 ghz C19 dbm nominal displayed average noise level (danl) (input terminated, sample or average detector, averaging type = log, 0 db input attenuation, if gain = high, 20 to 30 c) parentheses indicate typical performance preamplifer off preamplifer on 10 hz C95 dbm nominal 100 hz C110 dbm nominal 1 khz C115 dbm nominal 9 khz to 1 mhz (C125) dbm 1 to 10 mhz C14 4, (C14 8) dbm C154, (C158) dbm 10 mhz to 1.5 ghz C148, (C150) dbm C160, (C163) dbm 1.5 to 4.5 ghz C146, (C149) dbm C160, (C163) dbm 4.5 to 7 ghz C141, (C145) dbm C157, (C161) dbm 7 to 9.5 ghz C14 4, (C147) dbm C158, (C160) dbm 9.5 to 13 ghz C136, (C140) dbm C156, (C160) dbm 13 to 14.5 ghz C142, (C145) dbm C158, (C161) dbm 14.5 to 19.3 ghz C132, (C138) dbm C153, (C157) dbm 19.3 to 23 ghz C134, (C139) dbm C152, (C157) dbm 23 to 24 ghz C132, (C137) dbm C150, (C155) dbm 24 to 26.5 ghz C128, (C133) dbm C14 4, (C149) dbm spurious responses residual response (input terminated and 0 db attenuation) 200 khz to 26.5 ghz (swept) C90 dbm zero span or fft or other frequencies C100 dbm nominal tuned frequency (f) excitation freq mixer level response image responses (second mixer) 10 mhz to 7.5 ghz f + 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) 7.5 ghz to 19.3 ghz f - 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) 19.3 ghz to 21 ghz f + 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) 21 ghz to 22.8 ghz f - 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) 22.8 ghz to 24.2 ghz f + 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) 24.2 ghz to 26.5 ghz f - 1645 mhz ?10 dbm ?70 dbc (?80 dbc typical) lo-related spurious 10 mhz to 26.5 ghz ?10 dbm ?64 dbc typical other spurious responses mixer level response if feedthrough ?10 dbm ?75 dbc (?80 dbc typical) first rf order (f 10 mhz from carrier) ?10 dbm ?70 dbc (?80 dbc nominal) high rf order (f 10 mhz from carrier) ?30 dbm ?70 dbc (?80 dbc nominal) second harmonic distortion (shi) source frequency shi (nominal) 10 mhz to 3.75 ghz +50 dbm 3.75 to 13.25 ghz +62 dbm dynamic range specifcations
10 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet third-order intermodulation distortion (toi) parentheses indicate typical performance preamp off (two C20 dbm tones at input mixer spaced by 100 khz, 0 db attenuation, 20 to 30 c 10 mhz to 2 ghz +12 dbm, (+16) dbm 2 to 3 ghz +12 dbm, (+17) dbm 3 to 7.5 ghz +12 dbm, (+16) dbm 7.5 to 13.6 ghz +11 dbm, (+15) dbm 13.6 to 26.5 ghz +10 dbm, (+14) dbm option p03/p07/p13/p26 preamp on (two C45 dbm tones at the preamp input, spaced by 100 khz, 0 db attenuation, 20 to 30 c) 10 mhz to 26.5 ghz C8 dbm nominal figure 1. nominal dynamic range for band 0, for second and third order distortion, 10 mhz to 3 ghz figure 2. nominal dynamic range, for second and third order distortion, 3 ghz to 26.5 ghz dynamic range specifcations (continued)
11 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet phase noise offset specifcation typical noise sidebands (20 to 30 c, cf = 1 ghz) 100 hz C90 dbc/hz nominal 1 khz C102 dbc/hz C105 dbc/hz 10 khz C106 dbc/hz C110 dbc/hz 100 khz C108 dbc/hz C110 dbc/hz 1 mhz C130 dbc/hz C132 dbc/hz 10 mhz C145 dbc/hz nominal dynamic range specifcations (continued) figure 3. nominal phase noise at different center frequencies
12 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet channel power amplitude accuracy, w-cdma or is95 (20 to 30 c, attenuation = 10 db) 1.23 db ( 0.62 db 95th percentile) occupied bandwidth frequency accuracy [span/1000] nominal adjacent channel power accuracy, w-cdma (aclr) (at specifc mixer levels and aclr ranges) adjacent alternate ms 0.21 db 0.25 db bts 0.89 db 0.67 db dynamic range (typical) without noise correction C68 db C72 db with noise correction C73 db C78 db offset channel pairs measured 1 to 6 multiple number of carriers measured up to 12 power statistics ccdf histogram resolution 0.01 db harmonic distortion maximum harmonic number 10th results fundamental power (dbm), relative harmonics power (dbc), total harmonic distortion in % intermod (toi) measure the third-order products and intercepts from two tones burst power methods power above threshold, power within burst width results single burst output power, average output power, maximum power, minimum power within burst, burst width spurious emission w-cdma (1 to 3.0 ghz) table-driven spurious signals; search across regions dynamic range 75.0 db (79.1 db typical) absolute sensitivity ?82.5 dbm (?86.5 dbm typical) spectrum emission mask (sem) cdma2000 ? (750 khz offset) relative dynamic range (30 khz rbw) 73.0 db (78.1 db typical) absolute sensitivity ?99.7 dbm (?102.7 dbm typical) relative accuracy 0.11 db 3gpp w-cdma (2.515 mhz offset) relative dynamic range (30 khz rbw) 7 7.5 db (82.4 db typical) absolute sensitivity ?97.7 dbm (?101.7 dbm typical) relative accuracy 0.15 db powersuite measurement specifcations
13 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet output frequency frequency range option t03 2 mhz to 3 ghz option t07 2 mhz to 7.5 ghz option t13 2 mhz to 13.6 ghz option t26 2 mhz to 26.5 ghz resolution 10 hz output power level range 2 mhz to 10 ghz 35 to 0 dbm 10 to 20 ghz 35 to 5 dbm 20 to 26.5 ghz 40 to 12 dbm resolution 0.1 db absolute accuracy (at 50 mhz, C15 dbm) 1.0 db typical 0.3 db nominal output fatness (referenced to 50 mhz, C15 dbm) 95th percentile ( 2) 2 mhz to 7.5 ghz 1.0 db 0.3 db 7.5 to 13.6 ghz 1.2 db 0.3 db 13.6 to 23 ghz 1.8 db 0.6 db 23 to 26 ghz 2.5 db 1.2 db 26 to 26.5 ghz 2.3 db level accuracy nominal 2 mhz to 7.5 ghz 0.8 db 7.5 to 13.6 ghz 0.9 db 13.6 to 23ghz 1.5 db 23 to 26 ghz 1.8 db 26 to 26.5ghz 2.9 db output power sweep range 2 mhz to 10 ghz 35 to 0 dbm 10 to 20 ghz 35 to 5 dbm 20 to 26.5 ghz 40 to 12 dbm resolution 0.1 db maximum safe reverse level average total power + 30 dbm (1 w) ac coupled 50 vdc phase noise noise sidebands (center frequency = 1 ghz) offset typical 10 khz 98 dbc/hz 100 khz 105 dbc/hz 1 mhz 122 dbc/hz dynamic range maximum output power C displayed average 110 dbc nominal output vswr 2 mhz to 7 ghz < 1.7:1 nominal 7 to 23 ghz < 2.5:1 nominal 23 to 26.5 ghz < 3.5:1 nominal tracking generator specifcations
14 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet general specifcations temperature range operating 0 to 55 c storage C40 to 70 c emc complies with european emc directive 2004/108/ec C iec/en 61326-1 C cispr pub 11 group 1, class a C as/nzs cispr 11 C ices/nmb-001 this ism device complies with canadian ices-001 cet appareil ism est conforme la norme nmb-001 du canada environmental stress samples of this product have been type tested in accordance with the keysight environmental test manual and verifed to be robust against the environ - mental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with iec 60068-2 and levels are similar to milprf-28800f class 3. power requirement power drawn from chassis 65 w weight net 1.9 kg (4.2 lbs) shipping 4.2 kg (9.3 lbs) dimensions height 132 mm (5.2 in) width 82 mm (3.2 in) length 202 mm (8.0 in) calibration cycle the recommended calibration cycle is one year; calibration services are available through keysight service centers inputs and outputs rf input connector 3.5mm female, 50 ? nominal rf output connector 3.5mm female, 50 ? nominal 10 mhz in connector smb male, 50 ? nominal 10 mhz out connector smb male, 50 ? nominal trigger in connector smb male, 10 k? nominal trigger out connector smb male, 50 ? nominal analog out connector smb male, 50 ? nominal noise source for use with keysight 346 and sns series noise sources
15 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet i/q analyzer frequency frequency span standard instrument 10 hz to 10 mhz option b25 10 hz to 25 mhz resolution bandwidth (spectrum measurement) range overall 100 mhz to 3 mhz span = 1 mhz 50 hz to 1 mhz span = 10 khz 1 hz to 10 khz span = 100 hz 100 mhz to 100 hz window shapes flat top, uniform, hanning, gaussian, blackman, blackman-harris, kaiser bessel (k-b 70 db, k-b 90 db and k-b 110 db) analysis bandwidth standard instrument 10 hz to 10 mhz option b25 10 hz to 25 mhz if frequency response (standard 10 mhz if path) if frequency response (demodulation and fft response relative to the center frequency, 20 to 30 c) center frequency (ghz) span (mhz) max. error rms (nominal) 3.0 10 0.40 db 0.03 db 3.0 < f 26.5 10 0.10 db if phase linearity (deviation from mean phase linearity, nominal) center frequency (ghz) span (mhz) peak-to-peak rms 3.0 10 0.5 0.2 3.0 < f 7.5 10 0.5 0.4 data acquisition (standard 10 mhz if path) time record length 4,000,000 iq sample pairs sample rate 90 msa/s adc resolution 16 bits option b25 25 mhz analysis bandwidth if frequency response (demodulation and fft response relative to the center frequency, 20 to 30 c) center frequency (ghz) span (mhz) max. error rms (nominal) 3.0 10 to 25 0.45 db 0.03 db 3.0 < f 26.5 10 to 25 0.20 db if phase linearity (deviation from mean phase linearity, nominal) center frequency (ghz) span (mhz) peak-to-peak rms 0.02 f < 3.0 10 to 25 1 0.3 3.0 < f 7.5 10 to 25 1 0.5 data acquisition (b25 if path) time record length iq analyzer 4,000,000 iq sample pairs sample rate 90 msa/s adc resolution 16 bits system requirements operating system windows 7 (32 & 64 bit) processor speed 1.86 ghz minimum 2.4 ghz recommended available memory 4 gb minimum 8 gb recommended available disk space 4 gb video support for directx 9 graphics with 128 mb graphics recommended (supervga supported) browser microsoft internet explorer 7.0 or greater
16 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet instrument connection software keysight io library the io library suite offers a single entry point for connection to the most common instruments including axie, pxi, gpib, usb, ethernet/lan, rs-232, and vxi test instruments from keysight and other vendors. it automatically discovers interfaces, chassis, and instruments. the graphical user interface allows you to search for, verify, and update ivi instrument and soft front panel drivers for modular and traditional instruments. the io suite safely installs in side-by-side mode with ni i/o software. free software download at www.keysight.com/fnd/ iosuite module setup and usage keysight soft |front panel the cxa-m includes a soft front panel (sfp), a software-based graphical user interface (gui) which enables the instruments capabilities from your pc. included on cd-rom shipped with module or online module management keysight connection expert connection expert is the graphical user interface included in the io libraries suite that allows you to search for, verify and update ivi instrument and soft front panel drivers for modular and traditional instruments free software download at www.keysight.com/fnd/ iosuite programming driver development environments lvi-com visual studio (vb .net, c#, c/c++), vee, labview, labwindows/ cvi, matlab included on cd-rom shipped with module. ivi-c labview matlab programming assistance command expert assists in fnding the right instrument commands and setting correct parameters. a simple interface includes documentation, examples, syntax checking, command execution, and debug tools to build sequences for integration in excel, matlab, visual studio, and vee. free software download at www.keysight.com/fnd/ commandexpert signal analysis software x-series measurement applications provides measurements for analog demodulation, noise fgure, phase noise, vector signal analysis, and others. licensed software. for more information, visit www.keysight.com/ fnd/x-series_apps 89600 vsa 89600 vsa software sees through the complexity of emerging and existing industry standards, serving as your window into complex signal interactions. licensed software. for more information, visit www.keysight.com/fnd/vsa systemvue systemvue is a system-level eda platform for designing communi - cations and defense systems. used with the m9290a, systemvue enables you to create model-based design validation tests to ensure consistency from design to manufacturing. licensed software. for more information, visit www.keysight.com/fnd/ systemvue software
17 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet related literature literature pub number m9290a cxa-m pxie signal analyzer - product fact sheet 5992-0044en m9290a cxa-m pxie signal analyzer - confguration guide 5992-0193en m9018a pxie 18 slot chassis - data sheet 5990-6583EN m9037a pxie high performance embedded controller - data sheet 5991-3661en m9036a pxie embedded controller - data sheet 5990-8465en web for more information or literature resources please visit the web: product page: www.keysight.com/find/m9290a x-series measurement applications: www.keysight.com/find/x-series_apps x-series signal analyzers: www.keysight.com/find/x-series
mykeysight www.keysight.com/find/mykeysight a personalized view into the information most relevant to you. http://www.keysight.com/find/emt_product_registration register your products to get up-to-date product information and find warranty information. keysight services www.keysight.com/find/service keysight services can help from acquisition to renewal across your instruments lifecycle. our comprehensive service offeringsone- stop calibration, repair, asset management, technology refresh, consulting, training and morehelps you improve product quality and lower costs. keysight assurance plans www.keysight.com/find/assuranceplans up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. keysight channel partners www.keysight.com/find/channelpartners get the best of both worlds: keysights measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/cxa-m 18 | keysight | m9290a cxa-m pxie x-series signal analyzer - data sheet for more information on keysight technologies products, applications or services, please contact your local keysight office. the complete list is available at: www.keysight.com/find/contactus americas canada (877) 894 4414 brazil 55 11 3351 7010 mexico 001 800 254 2440 united states (800) 829 4444 asia pacifc australia 1 800 629 485 china 800 810 0189 hong kong 800 938 693 india 1 800 11 2626 japan 0120 (421) 345 korea 080 769 0800 malaysia 1 800 888 848 singapore 1 800 375 8100 taiwan 0800 047 866 other ap countries (65) 6375 8100 europe & middle east austria 0800 001122 belgium 0800 58580 finland 0800 523252 france 0805 980333 germany 0800 6270999 ireland 1800 832700 israel 1 809 343051 italy 800 599100 luxembourg +32 800 58580 netherlands 0800 0233200 russia 8800 5009286 spain 800 000154 sweden 0200 882255 switzerland 0800 805353 opt. 1 (de) opt. 2 (fr) opt. 3 (it) united kingdom 0800 0260637 for other unlisted countries: www.keysight.com/find/contactus (bp-9-7-17) dekra certified iso9001 quality management system www.keysight.com/go/quality keysight technologies, inc. dekra certified iso 9001:2015 quality management system evolving since 1939 our unique combination of hardware, software, services, and people can help you reach your next breakthrough. we are unlocking the future of technology. from hewlett-packard to agilent to keysight. this information is subject to change without notice. ? keysight technologies, 2014 - 2017 published in usa, october 17, 2017 5992-0032en www.keysight.com


▲Up To Search▲   

 
Price & Availability of 5990-6583EN

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X